HDT/VICAT Testing machine

Product Images

  • HDT/VICAT Testing machine

Product Property

Product Status : New
Sample Available : no
Shipment Terms : fob,cfr,cif,Negotiable
Payment Mode : T/T

Description

Functions

This series of HDT/VICAT testing machine is mainly used to determine temperature of deflection under load and Vicat softening temperature of thermoplastic plastics, pipes and fittings.

Standards

ISO2507, ISO75, ISO306, ASTM D648, ASTM D1525

GB/T 8802, GB/T 1633, GB/T 1634

Features

Ø HDT and VICAT test can be conducted at the same time

Ø Switzerland made high precision micrometer provides high accurate displacement measurement with high stability

Ø Up to six test stations, maximally up to six deformation transducers, and one temperature transducer

Ø All test stations are motor-driven, raised from the bath by the touch of a button on the control panel, allowing easy placement of the test specimen

Ø Water cooling and nature cooling are available

Ø Professional test software with friendly interface, easy to operate

Ø Once the software is started, the test is completely automatic and is performed according to the user-defined program. Upon test completion, it will automatically cool the bath to the starting temperature in preparation for subsequent tests. Using tap water, the temperature of the silicon oil can be reduced from a maximum of 300°C to 20°C.

Loading nose for HDT

Needle for VICAT

Perform HDT and VICAT in the same station.

Only change loading nose for HDT test, and needle for VICAT test.

Close view of single station

Test software

Ø

HDT test curve

VICAT test curve

Friendly and easy operation interface

Ø Test resume function when a sudden shutdown happens

Ø Powerful monitor function on test curve

Ø Test curve can be real-time zoomed in

Ø Multi-station test curves can be generated on a single report

Ø Test report can be printed out

Ø Test report can be exported to EXCEL for review and print


Specifications

Model

HVT302

HVT302

HVT302

Type

A

B

C

Test station

3

4

6

Immersion bath

1

1

1

Heat transfer medium

Methyl Silicon oil or transformer oil

Temperature sensor

1

1

1

Temperature range

Ambient (20)~300

Temperature accuracy

±0.5

Heating speed

Speed A

5±0.5/6min

Speed B

12±1/6min

Micrometer

3

4

6

Deformation measurement range

±5mm

Deformation resolution

0.001mm

Deformation measurement accuracy

0.003μm

Vicat Loading capacity:

GA=10N±0.2N, GB=50N±1N

HDT test

Method A: USE FIBRE STRESS 1.80Mpa

Method B: USE FIBRE STRESS 0.45Mpa

Method C: USE FIBRE STRESS 8.00Mpa

Maximum heating power

4500W

5000W

Power requirements

AC380V±10% 50HZ 15A

Dimension (LxWxH)

645x670x1450mm

835x670x1450mm

Weight

200kg

280kg

Cooling

Nature cooling or water cooling

Standard accessories:

Model

HVT302

HVT302

HVT302

Type

A

B

C

Main frame

1 set

1 set

1 set

Test stations

3

4

6

Swiss Micrometers

3 pieces

3 pieces

3 pieces

Temperature sensors

1 piece

1 piece

1 piece

Loading nose for HDT

1 set

1 set

1 set

Needle for VICAT

1 set

1 set

1 set

Weights (50N)

1 set

1 set

1 set

Test software

1 set

1 set

1 set